RT-V

Cigarette & Filter Rod Multifunction Test Bench

Measures weight, circumference (roundness), length, draw resistance, ventilation and hardness of cigarettes and filter rods. Modular design in which every test unit measures independently and simultaneously.

← Back to Cigarette & Filter Rod Physical Testing

Measured Parameters

  • Weight
  • Circumference
  • Roundness
  • Length
  • Draw resistance / pressure drop
  • Ventilation rate
  • Hardness

Features

  • Modular drawer structure: combine units as needed and switch individual test items on or off
  • Automatic roller hopper switches between king-size, slim and superslim grades
  • Weighing unit with imported high-precision balance (EMI-resistant), automatic calibration and tare
  • Circumference unit with user-defined measuring position; can test transparent filter rods
  • Draw resistance unit with valve block, wedge seal, temperature/humidity monitoring and automatic calibration with standard rods; test line pressure drop <10 Pa
  • Air-bearing hardness unit with non-contact laser measurement and automatic sample centering
  • 15-inch industrial tablet control; thermal printing, network and USB data upload; external alarm light; remote diagnosis and maintenance

Options

  • Sorting unit: sorts mixed samples by preset parameters
  • Multi-hopper system: samples of several grades queue for testing
  • Built-in UPS: keeps the instrument powered during outages
  • FOAS filter rod online automatic sampling system
  • COAS cigarette online automatic sampling system
  • Moisture and density testing system
  • Custom configurations for special parameters

Technical Specifications

ModelRT-V
Hopper capacity70 pcs (at 7.8 mm diameter)
Operating speedUp to 6 pcs/min
Sample collectionUnordered, more than 200 pcs
Hardness test point to end face0–75 mm
Circumference test point to end face47–119 mm
Constant air flowJewel CFO critical flow orifice, 17.5±0.1 ml/s
Overall sizeLow cabinet 600×600×1400 mm; high cabinet 600×600×1600 mm (excluding alarm light and computer)
ItemRangeResolutionAccuracy
Weight0–200 g0.001 g±0.001 g
Circumference15.5–28.5 mm0.001 mm±0.002 mm
Length55–150 mm0.001 mm±0.01 mm
Draw resistance0–10 kPa1 Pa±0.5% FS
Ventilation rate0–100%0.1%±1.0% FS
Pre-load0.1 N (adjustable)0.1 g±1 g
Load300 g (GB)0.1 g±1 g
Time15.0 s (adjustable)0.1 s±0.1 s
Displacement0–10 mm0.005 mm±0.01 mm
Hardness0–100%0.1%±0.1%

Operating Environment

Ambient humidity40%–65% RH (recommended (60±5)% RH)
Ambient temperature+10 ~ +35 °C (recommended (22±2) °C)
Power supplyAC 220 V ±10%, 50 Hz ±5%
Air supply0.5–0.8 MPa, ≥80 L/min, free of water, oil and dust
OtherKeep the instrument out of direct sunlight and other strong light

Standards

  • GB/T 22838.2-2009
  • GB/T 22838.3-2009
  • GB/T 22838.4-2009
  • GB/T 22838.5-2009
  • GB/T 22838.6-2009

Applications

  • Laboratory testing
  • Production-floor sampling

Related Equipment

Send an Inquiry

Tell us the products, specifications and quantities you need, and we will reply promptly.